1. 서론
This manual provides detailed instructions for the operation and maintenance of the OWON SDS1102 Digital Storage Oscilloscope. The SDS1102 is a 2-channel desktop oscilloscope offering a 100MHz bandwidth and a 1GS/s real-time sample rate. It features a 7-inch high-resolution LCD and supports SCPI commands and LabVIEW for advanced control and data acquisition. Please read this manual thoroughly before using the device to ensure proper and safe operation.
2. 안전 정보
To prevent electric shock or personal injury, and to avoid damage to the oscilloscope or connected equipment, observe the following safety precautions:
- Ground the Product: The oscilloscope must be properly grounded.
- Connect and Disconnect Probes Properly: Connect the probe output to the oscilloscope before connecting the probe to the circuit under test. Disconnect the probe from the circuit under test before disconnecting it from the oscilloscope.
- 모든 단자 정격을 준수하세요: To prevent fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for more information about ratings before making connections to the product.
- 커버 없이 작동하지 마십시오: 커버나 패널을 제거한 상태로 제품을 작동하지 마십시오.
- Use Proper Fuse: Use only the specified fuse type and rating.
- Avoid Exposed Circuitry: Do not touch exposed connections or components when power is present.
- 젖은/D에서 작동하지 마십시오amp 정황: Do not operate the product in wet or damp 정황.
- Do Not Operate in Explosive Atmosphere: 폭발성 환경에서 제품을 작동하지 마십시오.
- 제품 표면을 깨끗하고 건조하게 유지하세요: Keep product surfaces clean and dry to prevent contamination.
3. 패키지 내용
Verify that all items listed below are included in your package. If any items are missing or damaged, please contact your distributor.

그림 3.1: OWON SDS1102 Digital Storage Oscilloscope with included accessories. The image displays the main oscilloscope unit, two oscilloscope probes, a power cord, a USB cable, and a software CD.
- OWON SDS1102 Digital Storage Oscilloscope Unit
- Oscilloscope Probes (2 units)
- 전원 코드
- USB 케이블
- Quick Guide / Software CD (may vary by region)
4. 제품 오버view
The OWON SDS1102 is designed for precise waveform analysis with its intuitive interface and robust features.
4.1 주요 특징
- 100MHz Bandwidth, 2 Channels
- 1GS/s Real-time Samp르 레이트
- 7-inch (800x480 pixels) High-Resolution Color LCD
- Supports SCPI Remote Commands and LabVIEW
- USB Host for U-disk storage and firmware upgrades
- Built-in 6-bit hardware frequency meter
- 30 types of automatic measurements
- Ultra-thin body design (70mm thickness, 1.1kg weight)
4.2 전면 패널

그림 4.1: 앞쪽 view of the OWON SDS1102 Oscilloscope highlighting its display and controls. The image shows the 7-inch LCD screen, control knobs, function buttons, and input connectors for channels 1 and 2. Key specifications like 1GS/s Sample rate, 10K Record Length, 2-Channel, 100 MHz Bandwidth, and 7 inch high resolution LCD are overlaid.
The front panel includes the 7-inch LCD display, various control buttons, rotary knobs for horizontal and vertical adjustments, trigger controls, and the input connectors for Channel 1 (CH1) and Channel 2 (CH2). A USB host port is conveniently located on the front for data storage and firmware updates.

그림 4.2: Detail of the front-sided USB port. This image emphasizes the USB port on the front panel, indicating its use for saving data (waveform records) and its capacity for up to 10K samp링 포인트.
4.3 물리적 치수

그림 4.3: Physical dimensions and weight of the SDS1102. The image shows the oscilloscope with its dimensions: 11.85 inches (301mm) width, 5.98 inches (152mm) height, and 2.75 inches (70mm) thickness. The total weight is 1.1 kg. It also highlights the ultra-thin design and the side-mounted power socket.
The SDS1102 features an ultra-thin design, measuring 301mm (width) x 152mm (height) x 70mm (thickness) and weighing approximately 1.1 kg, making it highly portable and space-efficient for desktop use.
4.4 디스플레이 인터페이스

그림 4.4: The 7-inch color LCD display. This image focuses on the oscilloscope's 7-inch color LCD screen, emphasizing its 800 x 480 pixel resolution for clear and concise waveform readings.
The 7-inch (800x480 pixels) color LCD provides a clear and detailed display of waveforms and measurement parameters. The display grid is 8x12, offering ample space for waveform visualization.
5. 설정
5.1 전원 연결
- 오실로스코프의 전원 스위치가 OFF 위치에 있는지 확인하세요.
- Connect the provided power cord to the power input socket on the side of the oscilloscope.
- Plug the other end of the power cord into a grounded AC power outlet (100V-240V AC, 50/60Hz).
5.2 프로브 연결
- Connect the BNC connector of the oscilloscope probe to the CH1 or CH2 input connector on the front panel.
- 프로브가 제자리에 단단히 고정되었는지 확인하십시오.
- 프로브 접지 클립을 회로의 접지 지점에 연결하십시오.
- Set the probe attenuation switch (usually on the probe itself) to the desired setting (e.g., 1X or 10X). Remember to match this setting in the oscilloscope's channel menu for accurate readings.
5.3 초기 전원 켜기
After connecting power and probes, press the power button on the front panel to turn on the oscilloscope. The device will perform a self-test and display the main waveform interface.
6. 사용 설명서
6.1 기본 파형 표시
Use the vertical (VOLTS/DIV) and horizontal (SEC/DIV) knobs to adjust the scale of the displayed waveform. The position knobs allow you to move the waveform vertically and horizontally on the screen.
6.2 트리거링
The trigger system stabilizes repetitive waveforms and captures single-shot events. Adjust the trigger level and mode using the dedicated trigger controls on the front panel to achieve a stable display.
6.3 Cursor Measurements
The SDS1102 supports cursor measurements for precise analysis of waveform parameters.

그림 6.1: Example of cursor measurements. This image shows the oscilloscope display with two vertical cursors (A and B) placed on a sine wave. The measurement results, including time difference (Δx) and voltage difference (Δy), are displayed on the screen.
To use cursor measurements:
- 를 누르세요 커서 단추.
- Select the desired cursor type (e.g., Voltage, Time, Track) from the menu.
- Use the multipurpose knob to position the cursors on the waveform.
- The measured values (e.g., ΔV, ΔT) will be displayed on the screen.
6.4 Automatic Measurements
The oscilloscope offers 30 types of automatic measurements, providing quick and accurate parameter readings.

그림 6.2: Automatic measurement results. This image displays a comprehensive list of automatic measurements for a waveform, including Max, Min, PK-PK, Vpp, Mean, RMS, Cycle RMS, Rise Time, Fall Time, Frequency, Period, Duty Cycle, and more.
To access automatic measurements:
- 를 누르세요 측정하다 단추.
- Select the desired measurement type from the menu.
- The results will be displayed on the screen.
6.5 파형 저장 및 불러오기
The SDS1102 allows for storing and recalling waveforms. You can save waveforms to the internal memory or to a USB flash drive connected to the front-sided USB port.

그림 6.3: Waveform storage menu. This image shows the oscilloscope's menu for saving and recalling waveforms, with options to select different waveform slots (Wave 8 to Wave 15) and to save or recall the selected waveform.
6.6 PC Communication (SCPI and LabVIEW)
The SDS1102 supports communication with a host computer via USB, enabling remote control using SCPI commands and integration with LabVIEW for advanced data analysis and automation.

그림 6.4: Oscilloscope connected to a laptop for PC communication. The image shows the OWON SDS1102 oscilloscope connected to a laptop via a USB cable, demonstrating its capability for communication and remote control using SCPI and LabVIEW.
To establish PC communication:
- Install the necessary drivers and software (e.g., LabVIEW drivers, SCPI command interface) on your computer.
- Connect the oscilloscope to your computer using the provided USB cable.
- Refer to the OWON website or software documentation for detailed instructions on using SCPI commands or LabVIEW 완성.
7. 유지관리
7.1 Cleaning the Oscilloscope
오실로스코프의 성능과 외관을 유지하려면 다음 사항을 준수하십시오.
- 청소하기 전에 전원 코드를 뽑으세요.
- 부드러운 천을 사용하세요ampened with mild detergent and water to clean the exterior surfaces.
- 플라스틱 부품을 손상시킬 수 있는 연마성 세척제나 용제를 사용하지 마십시오.
- 기기에 액체가 들어가지 않도록 하십시오.
7.2 펌웨어 업데이트
Firmware updates may be released periodically to improve performance or add new features. These updates can typically be performed via the front-sided USB port using a USB flash drive.
- Visit the official OWON website for the latest firmware versions and detailed update instructions.
- 펌웨어 다운로드 file USB 플래시 드라이브에.
- Follow the on-screen prompts or specific instructions provided with the firmware update.
8. 문제 해결
이 섹션에서는 일반적으로 발생할 수 있는 문제에 대한 해결책을 제공합니다.
| 문제 | 가능한 원인 | 해결책 |
|---|---|---|
| 파형이 표시되지 않습니다 |
|
|
| 불안정한 파형 |
|
|
| 전원이 없습니다 |
|
|
If the problem persists after attempting these solutions, please contact OWON customer support.
9. 사양
| 매개변수 | 세부 사항 |
|---|---|
| 모델 | SDS1102 |
| 대역폭 | 100MHz |
| 채널 | 2 |
| Samp르 레이트 | 1 GS/s (실시간) |
| 표시하다 | 7-inch Color LCD, 800 x 480 pixels |
| 레코드 길이 | 10K points |
| 인터페이스 | USB Host, USB Device (supports SCPI, LabVIEW) |
| 자동 측정 | 30가지 유형 |
| 전원 공급 장치 | 100V-240V AC, 50/60Hz |
| 치수(길이 x 높이 x 너비) | 301 x 152 x 70mm(11.85 x 5.98 x 2.75인치) |
| 무게 | 1.1kg(2.43파운드) |