1. 서론
Thank you for choosing the WEST TUNE Laboratory Micro Pipette Explorer Series. This manual provides essential information for the safe and efficient operation, maintenance, and troubleshooting of your new pipette. Designed for precision and ease of use, the Explorer Series micropipettes are ideal for a wide range of laboratory, industrial, food, cosmetic, and chemical samp링 응용 프로그램.

2. 주요 특징
- Accurate Liquid Extraction: Volume range from 0.1µL to 10mL, with precise volume adjustment via a rotating piston button and clear digital display.
- 보증 된 품질 : Calibrated to ISO8655 standards, factory-adjusted for high accuracy with distilled or deionized water. Each unit includes an individual test certificate.
- 인체공학적 디자인: Lightweight and designed for comfortable, flexible manual operation.
- Efficient Liquid Handling: Features a blow-out function for complete liquid discharge, ensuring no residue.
- Durable & Maintainable: Constructed from chemically resistant materials, with removable tips. Half-autoclavable for sterilization at 121°C high temperature and high pressure.
- 다양한 응용 프로그램: Suitable for various settings including laboratories, industrial processes, food testing, cosmetic experiments, pool water analysis, and general chemical samp링.

3. 설치 및 조립
3.1 초기 검사
Upon receiving your pipette, inspect the packaging and contents for any signs of damage. Ensure all components listed in the 'Package Contents' section are present.
3.2 피펫 팁 부착하기
- Select the appropriate pipette tip for your desired volume range.
- Firmly attach the tip to the tip ejector collar of the pipette. Ensure a secure seal to prevent leaks and ensure accurate pipetting.
3.3 볼륨 조절
The volume is adjusted by rotating the piston button at the top of the pipette. The digital display will show the selected volume. Rotate clockwise to increase volume and counter-clockwise to decrease. Always ensure the desired volume is clearly visible in the digital window.

3.4 Disassembly and Reassembly for Maintenance
For detailed instructions on disassembling and reassembling the pipette for cleaning or maintenance, please refer to the video below. This process involves using the provided calibration tool and dismantling device to separate the tip collar, tip ejector, handle connector, counter component, color cap, screw components, latch, button, spring, spring pad, and adjust screw nut.
4. 사용 설명서
4.1 Liquid Suction
- Attach a clean, appropriate pipette tip.
- Set the desired volume using the piston button.
- Press the piston button down to the first stop.
- Immerse the tip into the liquid to be aspirated.
- Slowly release the piston button to draw the liquid into the tip.
4.2 Liquid Discharge
- 팁을 수용 용기의 내벽에 약간 비스듬한 각도로 대십시오.
- Press the piston button down to the first stop to dispense the liquid.
- To ensure complete discharge and prevent residue, continue pressing the piston button down to the second stop (blow-out function).
- Hold for a moment, then release the piston button.

4.3 팁 배출
After dispensing the liquid, use the tip ejector to safely remove the used tip. Press the tip ejector button to release the tip into a waste container.

5. 유지관리
5.1 청소
Regular cleaning is crucial for maintaining the accuracy and longevity of your pipette. Disassemble the lower part of the pipette (refer to Video 1 in Section 3.4) and clean components with a mild detergent solution. Rinse thoroughly with distilled water and allow to air dry completely before reassembly.
5.2 살균
The WEST TUNE Explorer Series pipettes are half-autoclavable. This means the lower part of the pipette (tip ejector, tip ejector collar, and piston component) can be sterilized at 121°C (250°F) and 15 psi for 20 minutes. Ensure all components are completely dry before reassembly.

5.3 재보정
If you suspect the pipette's volume is inaccurate, it can be recalibrated. This process typically requires an analytical balance and specific calibration weights. Refer to the detailed instructions in the included user manual or consult a qualified technician for recalibration procedures.
6. 문제 해결
Most common issues can be resolved with proper maintenance and technique. Here are some common problems and their solutions:
- Inaccurate Volume:
- Ensure the pipette tip is securely attached and not leaking.
- Verify the digital display shows the correct volume setting.
- Check for any blockages or damage in the tip or pipette shaft.
- If the issue persists, recalibrate the pipette (refer to Section 5.3).
- Liquid Dripping from Tip:
- Ensure the tip is properly seated and forms a tight seal.
- Check for any damage to the tip or the pipette's tip ejector collar.
- Verify the piston is moving smoothly without resistance.
- Difficulty in Aspiration/Dispensing:
- Ensure the tip is not clogged.
- Check for proper piston movement; lubricate if necessary (use calibration oil).
- Verify the liquid viscosity is appropriate for the pipette's range.
7. 사양
재료: 플라스틱
분류: 피펫
고위험 화학물질: 없음
볼륨 범위: 0.1µL to 10mL (depending on model)

| 볼륨 범위 | 증가 | Test Volume (µL) | Accuracy Error | 정밀도 오류 | ||
|---|---|---|---|---|---|---|
| % | μL | % | μL | |||
| 0.1-2.5μL | 0.05µL | 2.5 1.25 0.25 | 2.50% 3.00% 12.00% | 0.0625 0.0375 0.03 | 2.00% 3.00% 6.00% | 0.05 0.0375 0.015 |
| 0.5-10μL | 0.1µL | 10 5 1 | 1.00% 1.50% 2.80% | 0.1 0.075 0.028 | 0.80% 1.50% 0.028 | 0.08 0.075 0.015 |
| 2-20μL | 0.5µL | 20 10 2 | 0.90% 1.20% 3.00% | 0.18 0.12 0.06 | 0.40% 1.00% 2.00% | 0.08 0.1 0.04 |
| 5-50μL | 0.5µL | 50 25 5 | 0.60% 0.90% 2.00% | 0.3 0.225 0.1 | 0.30% 0.60% 2.00% | 0.15 0.15 0.1 |
| 10-100μL | 1µL | 100 50 10 | 0.80% 1.00% 3.00% | 0.8 0.5 0.3 | 0.15% 0.40% 1.50% | 0.15 0.2 0.15 |
| 20-200μL | 1µL | 200 100 20 | 0.60% 0.80% 3.00% | 1.2 0.8 0.6 | 0.15% 0.30% 1.00% | 0.3 0.3 0.2 |
| 50-200μL | 1µL | 200 100 50 | 0.60% 0.80% 1.00% | 1.2 0.8 0.5 | 0.15% 0.30% 0.40% | 0.3 0.3 0.2 |
| 100-1000μL | 5µL | 1000 500 100 | 0.60% 0.70% 2.00% | 6 3.5 2 | 0.20% 0.25% 0.70% | 2 1.25 0.7 |
| 200-1000μL | 5µL | 1000 500 200 | 0.60% 0.70% 0.90% | 6 3.5 1.8 | 0.20% 0.25% 0.30% | 2 1.25 0.6 |
| 1000-5000μL | 50µL | 5000 2500 1000 | 0.50% 0.60% 0.70% | 25 15 7 | 0.15% 0.30% 0.30% | 7.5 7.5 3 |
| 2~10ml | 0.1ml (XNUMXml) | 10ml 5ml 2ml | 0.60% 1.20% 3.00% | 60 60 60 | 0.20% 0.30% 0.60% | 20 15 12 |
8. 사용자 팁
- Always select the correct pipette model and tip for the specific volume range you intend to measure to ensure accuracy.
- When receiving your pipette, note that it may come with a single tip for initial use or testing. Additional tips are typically purchased separately.
- Handle the pipette with care to avoid dropping or impacting it, which can affect its calibration and performance.
- Regularly check the digital display for clarity and ensure the volume is set precisely before each use.
9. 보증 및 지원
Each WEST TUNE Micro Pipette is supplied with an individual test certificate, affirming its calibration to ISO8655 standards for precision and accuracy. Our products are designed for a long lifetime, exceeding 100,000 pipetting cycles, thanks to patent techniques and rigorous manufacturing experience. For any issues or support, please refer to the contact information provided with your purchase documentation. We value user feedback for continuous product improvement.
10. 패키지 내용
The standard package for the WEST TUNE Laboratory Micro Pipette Explorer Series typically includes:
- WEST TUNE Explorer Pipettor (selected volume range)
- 사용자 설명서
- Quality Control Certificate
- Accessory Supplied (e.g., calibration tool/wrench)
- Pipette Stand (may vary by package)
- Calibration Oil
- Pipette Tips (initial set, quantity may vary)